Spectrometer type and calibration requirement: The quoted model should be a laboratory benchtop model and suitable for Fe, Al & Cu base application. Instrument should be providing with factory Calibration for Fe-Orientation, Low Alloy, Cast Iron, Al- Global, Pur Aluminum and Cu-Orientation, Pure Copper, Cu/Zn Alloy as per mentioned element mentioned in the specification. (Annexure-1)
Optical System: Should have dual optical system. UV and Air Optical system.
Focal Length: 300 mm focal length for UV optics and 400 mm focal length for Air optics.
Type of Detectors: CMOS multi detector.
Source Parameters: Plasma Generator–Digital sample excitation system for high precision.
Wavelength Range: Minimum wavelength coverage from 120 nm – 670 nm or better.
Spark Stand: Spark Stand self-cooled with minimum argon gas consumption. Spark stand design should be such that samples of various kind of geometries can be tested.
Diagnosis System: Instrument should be equipped with Auto diagnostics indicating the status of parts in the system by different colour indication thereby enabling the operator to take adequate preventive measure to prevent breakdowns as Much as possible.
Application report: Bidder should submit manufacturer application report for Fe, Al, Cu base application.
Standardization: Instrument should have ICAL 2.0 facility for quick standardization of the instrument. Same should be mentioned in the offer and technical confirmation by OEM.